Column Compartment/oven with superior thermal performance, multistage (10 ramps) programmed temperature control function. (supported by “control system”)
Advanced built-in data acquisition system , supporting real time instrument status monitoring, detection signal acquisition and PC control
Column oven accommodates up to 3 chromatographic columns, and supports quick heat-up and rapid cool-down with automated back-door opening. 450℃ to 50℃ in 6 min at 25℃ ambient)
Flexible sample introduction system: 3 sample injectors could be installed and operated simultaneously with independent temperature control.
High sensibility and stability detector.
2 independent and analog signals output.
Clarity software, compatible with GLP/FDA-21 CFR Part11 requirements and regulations. (electronic records and signatures)
Special analyzers, like RGA, NGA, PINDA, etc. according to ASTM standards
Superior performance with advanced, microcomputer-based temperature control system
High temperature accuracy (optimum ± 0.02℃), high reliability, and anti-interference
Self-diagnosis/ self-protection function (overheat protection, power-off protection, etc.)
Intuitive display of timing program, detector status, measurement range, current setting, etc
Model# |
GC1290 |
GC1120 |
|
Description |
EPC |
|
|
Sample Introduction System |
sample injector and evaporation chamber |
||
Column Oven |
Temperature range |
Ambient temperature +7℃ ~ 400℃ (in 1℃ increment) |
|
Temperature accuracy |
± 0.02℃ |
||
Cooling time |
400℃ to 50℃ in 6 min at 25℃ ambient |
||
Programmed temperature setting |
0.1℃~40℃/min (in 1℃ increment) |
||
FID |
Detection limit |
≤3×10-12g/s (C16) |
|
Drift |
≤6×10-13A/h |
||
TCD |
Sensitivity |
≥5000mV.mL/mg |
|
Baseline noise |
≤20µV |
||
FPD |
Detection limit |
≤8×10-13g/s(P) ; ≤8×10-11g/s(S) |
|
Drift |
≤2×10-11A/30min |
||
Baseline noise |
≤5×10-12A |
||
ECD |
Detection limit |
≤2×10-13g/s (γ-666) |
|
Drift |
≤80µV/30min |
||
Baseline noise |
≤20µV |
||
Linear dynamic range |
≥103 |
||
NPD |
Detection limit |
≤5×10-13g/s(P) ; ≤5×10-12g/s(N) |