SUNNY HENGPING
INSTRUMENT
UV2000 Spectrophotometer
-
Single-point curve calibration and multi-point curve calibration (up to 12 points)
Display of fitting degree of curve fitting and concentration data with up to 50 sets of curve parameters storage
Display of concentration data with up to 200 sets of data storage
Concentration factor setting and concentration direct reading
Dark current calibration / wavelength calibration/system clock management
Optical System
|
Single beam, C-T type, 1200 lines/mm grating
|
Detector
|
Silicon photo diode
|
Light Source
|
Tungsten lamp (socket type, 20W / 12V, 2000 hours)
|
Deuterium lamp (1000 hours)
|
Display Mode
|
4.3” color LCD, touch screen
|
Spectral Bandwidth
|
4nm
|
Wavelength Range
|
200 - 1100nm
|
Wavelength Accuracy
|
±2.0 nm (Calibrate by system automatically )
|
Wavelength Repeatability
|
1.0nm
|
Stray Light
|
≤0.5%T (220Nal)
|
Photometric Range
|
T:-1.0 ~200.0%T ,A:-0.5 ~ 3.000Abs,C:0 ~ 9999
|
Photometric Accuracy
|
±0.5%T
|
Photometric Repeatability
|
0.2%T
|
Stability
|
dark current < 0.2%T, bright current < 0.5%T
|